ADVANCED RESEARCH SYSTEM

Advanced Research Systems (ARS)

provides a complete cryogenic solutionfor low temperature material characterization

Laboratory Cryogenic Systems    

ARS manufactured Pneumatic Drive Closed Cycle Cryocoolers are ideally suited for R&D applications where low temperature and low vibrations are important. ARS is uniquely qualified to design and manufacture customized cryogenic solutions as we can seamlessly integrate our cryocoolers into the final design.

Optical Cryostats

Low Vibration Cryostats

Top Loader Cryostats

Helium Flow Cryostats

UHV Cryostats

Cryogenic Probe Station

Mixed Gas Cryocooler

Closed Cycle Cryocooler Principles of Operation   

Click on the image for large size.

Closed Cycle Cryocooler Components

The major components of the closed cycle cryostat are the expander, compressor,  vacuum shroud, and radiation shield. The expander, commonly referred to as the coldhead or cold finger, is where the Gifford-McMahon refrigeration cycle takes pace.  It is connected to a compressor by two gas lines and an electrical power cable. One of the gas lines supplies high pressure helium gas to the expander, the other gas line returns low pressure helium gas from the expander.    The compressor provides the necessary helium gas flow rate at the high and low pressure for the expander to convert into the desired refrigeration capacity. The vacuum shroud surrounds the cold end of the expander in vacuum limiting the heat load on the expander caused by conduction and convection. The radiation shield is actively cooled by the first stage of the expander and insulates the second stage from the room temperature thermal radiation being emitted from the vacuum shroud.

In addition to these major components the closed cycle cryocooler is often accompanied by several support systems. Typically laboratory systems will have an instrumentation skirt, which provides a vacuum port and electrical feedthroughs, as well as a temperature controller to measure and adjust the sample temperature.  The system also requires electricity, cooling water for the compressor, and a vacuum pump for the sample space.

The Gifford-McMahon Refrigeration Cycle

The ARS closed cycle cryocoolers operate on a pneumatically driven Gifford-McMahon refrigeration cycle, often shortened to GM Cycle or GM cooler. The pneumatically driven GM cooler is  different mechanically driven GM coolers in that it uses an internal pressure differential to move the displacer instead of a mechanical piston, which results in smaller vibrations.

The refrigeration cycle of the ARS closed cycle cryostat starts with the rotation of the valve disk opening the high pressure path allowing the high pressure helium gas to pass through the regenerating material into the expansion space. Second, the pressure differential drives the piston "up" allowing the gas at the bottom to expand and cool. Third the rotation of the valve disk opens the low pressure path allowing the cold gas to flow through the regenerating material removing heat from the system. Finally the pressure differential returns the displacer to its original position completing the cycle.

Click on the image for large size.

FIG. 1. Block diagram of the internal structure of an ARS closed cycle cryocooler expander.

Click on the image for large size.

FIG. 2. Steps 1 & 2 of the GM cycle.

Click on the image for large size.